FAQ
General | |
---|---|
CMOS technology | 0.25μm |
Pixel size | 55μm x 55μm |
Pixel matrix | 256 x 256 |
Design | CERN |
Analog front end (pixel cell) | |
---|---|
Positive and negative signal polarity | |
Leakage current | -10nA to +20nA |
Time to peak | 100ns |
Noise | 100 e- rms |
Threshold variation (after tuning) | 35 e- rms |
Minimum operating treshold | 700 e- |
Digital part (pixel cell + periphery) | |
---|---|
1 counter / shift register | 14 bits (stops at 11810 counts) |
Periphery |
13 8-bit DACs to set voltages in the chip Serial raedout 1-bit LVDS Parallel readout 32-bit CMOS Readout time in serial mode (100 MHz clock): 9μs Readout time in parallel mode (100 MHz clock): 266μs |
Total analog power consumption (nominal conditions) | 440mW |
Total digital power consumption (@100MHz) | 450mW |
Features:
- Pixel size 55μm x 55μm
- 256 x 256 pixels
- Single energy threshold (adjustable per pixel with 4 bits)
- Three modes of operation: (1) single particle counting (2) Time over Threshold (TOT) and (3) Time of Arrival (TOA)
- Electron or hole collection -> compatible with various sensor materials
- Can be combined with Gas Gain Grid to readout electron deposition in a gas detector
- 3-side buttable
Applications:
- Adaptive optics and other visible or near visible light applications
- Astrophysics
- Background radiation monitoring
- Digital Autoradiography
- Dosimetry
- Education
- Electron microscopy
- Life Sciences
- Neutron imaging
- Various X-ray and gamma-ray imaging applications
- X-ray polarimetry measurements
General | |
---|---|
CMOS technology | 0.25μm |
Pixel size | 55μm x 55μm |
Pixel matrix | 256 x 256 |
Design | CERN |
Analog front end (pixel cell) | |
---|---|
Positive and negative signal polarity | |
Leakage current | -10nA to +20nA |
Time to peak | 100ns |
Noise | 100 e- rms |
Threshold variation (after tuning) | 35 e- rms |
Minimum operating treshold | 700 e- |
Digital part (pixel cell + periphery) | |
---|---|
1 counter / shift register | 14 bits (stops at 11810 counts) |
Periphery |
|
Total analog power consumption (nominal conditions) | 440mW |
Total digital power consumption (@100MHz) | 450mW |
Features:
- Pixel size 55μm x 55μm
- 256 x 256 pixels
- 2 thresholds for window discrimination (adjustable per pixel with 3 bits each)
- Single particle counting
- Electron or hole collection -> compatible with various sensor materials
- Can be combined with Gas Gain Grid to readout electron deposition in a gas detector
- 3-side buttable
Applications:
- Adaptive optics and other visible or near visible light applications
- Astrophysics
- Digital Autoradiography
- Education
- Electron microscopy
- Life Sciences
- Neutron imaging
- Various X-ray and gamma-ray imaging applications